Sample: 1.0 mL plasma and 50µL ISTD
Sorbent: SCX (1 mL, 100 mg, Varian)
Procedure: acidify sample with 50µL 5 N HCl, precondition SPE tube with 2 x 1 mL methanol and 2 x 1 mL deionized water, apply sample, rinse with 2 x 1 mL 0.01 M HCl, add 200µL 0.05 M sodium citrate pre-elution solution, elute with 750µL 3:97 ACN:0.1 M citric acid (pH 6.5)
Sample: 250µL sample and 50µL ISTD
Sorbent: SCX (3 mL, SPEC·PLUS, 15 mg, Ansys)
Automation Equipment: RapidTraceTM
SPE Workstation
Procedure: dilute sample with 0.2 mL deionized water, precondition SPE tube with 0.5 mL ACN and 0.5 mL 0.01 M HCl, apply sample, rinse with 2 mL 0.01 M HCl, elute with 250µL 10:90 ACN:0.1 M citric acid (pH 6.5)
Sample: 250µL sample and 50µL ISTD
Sorbent: SCX (MicroluteTM, 25 mg, Varian)
Procedure: dilute sample with 200µL deionized water, precondition SPE tube with >= 250µL ACN and >= 250µL 0.5 M HCl, apply sample, elute with 250µL 10:90 ACN:0.1 M citric acid (pH 6.5)
| . | nominal MTX concentration (ng/mL) | ||
|---|---|---|---|
| . | 550 | 90.0 | 15.0 |
| mean | 583 | 92.3 | 15.2 |
| std dev | 27.4 | 4.29 | 0.579 |
| rsd | 4.7% | 4.6% | 3.8% |
| accuracy | 106.0% | 102.5% | 101.2% |
| . | nominal MTX concentration (ng/mL) | ||
|---|---|---|---|
| . | 800 | 200 | 30.0 |
| mean | 815 | 211 | 30.8 |
| std dev | 65.7 | 22.0 | 2.43 |
| rsd | 8.1% | 10.4% | 7.9% |
| accuracy | 101.8% | 105.4% | 102.6% |
| . | nominal MTX concentration (ng/mL) | ||
|---|---|---|---|
| . | 800 | 200 | 30.0 |
| mean | 863 | 208 | 29.5 |
| std dev | 66.9 | 27.4 | 3.60 |
| rsd | 7.7% | 13.2% | 12.2% |
| accuracy | 107.8% | 103.8% | 98.3% |
* The Conventional Packed Bed Column Performs Well But Can Be More Time Consuming.
* The Automated Rapidtrace Extraction Takes About As Long As Conventional Packed Bed Sample Preparation, But It Frees Personnel For Other Tasks .
* The 96 Well Plate Approach Allows For Faster Sample Preparation, But When Done Manually It Is More Susceptible To Technique Variability (Ex. Recovery Problems, Etc). This Limitation Would Be Reduced By Automating The Reagent Delivery And By Better Control Of The Flow Rates.
* Reduced Elution Volume Afforded By SPEC Discs And Microlute Plates Gives A Concentration Enhancement For The "Elute And Shoot" Method, Improving The Signal To Noise Ratio Near The LLOQ And Allowing For A Reduced Sample Volume To Be Used.