BASi PWR-3 Compliance Voltage
In voltammetric measurements, current is measured as a function of the applied potential (Eapp). A potentiostat tries to
maintain Eapp between the working and reference electrodes and allows negligible current through the reference electrode. The
auxiliary electrode acts as a source, or sink, of electrons to balance the redox reaction occurring at the working electrode.
Therefore, the potential applied between the auxiliary and working electrodes (Eaux) must be sufficient both to drive the
appropriate electron transfer reaction at the auxiliary electrode and to compensate for the potential drop due to the solution
resistance between the auxiliary and reference electrodes. (This potential drop is given by Ohm's Law E = iRc, where i is the
current generated by the electrochemical reaction, and Rc is the solution resistance between the auxiliary and reference
electrodes.) The compliance voltage is the maximum potential that can be supplied between the working and auxiliary electrodes.

In a typical voltammetric experiment (e.g., cyclic voltammetry), µA currents are generated, and Rc is relatively
small. Therefore, the potential required between the working and auxiliary electrodes rarely exceeds a few volts. However, in
controlled potential electrolysis (bulk electrolysis) experiments, larger currents are generated (due to large surface area working
electrodes), and Rc is increased by the presence of the frit that is needed to separate the auxiliary and working
electrodes. In such experiments, a high compliance voltage may be required to maintain potential control (especially in non-aqueous
electrolytes). For example, if 0.2 A is generated and the resistance through the frit is 100 ohms, then a compliance voltage of at
least 20 V is required. If the compliance voltage of the potentiostat is not sufficient, then the actual potential applied between
the working and reference electrodes is lower than the required value and the electrolysis time will increase.
|