BASi Peroxidase Redox Polymer Wired Enzyme Electrode Kit

The electrochemical determination of hydrogen peroxide has been used to selectively measure acetylcholine, choline, glucose and
lactate in diverse biological samples. The analyte is isolated using a microbore or cartridge chromatography column. An
analyte-specific, oxidase enzyme bound to a post-column immobilized enzyme reactor (IMER) produces H2O2 which
is then detected downstream at the surface of a working electrode in the electrochemical detector.
Platinum
Platinum electrodes are frequently utilized because they are more selective for H2O2 than standard glassy
carbon electrodes. A glassy carbon electrode would have to operate at a much higher applied potential to detect peroxide and would
also respond to all of the other electroactive components in the sample. However, platinum requires long equilibration times after
cleaning or system changes.
A New Approach
The new wired enzyme electrode offers several advantages vs. the platinum electrode. It exhibits improved sensitivity, better
detection limits and improved operational stability. Initial stabilization of the background current is significantly faster than
for a conventional platinum electrode. A redox polymer film containing horseradish peroxidase (HRP) is coated on the surface of a
glassy carbon electrode. The redox polymer electrically "wires" the peroxidase enzyme to the electrode. It operates at a
lower applied potential than a platinum electrode (+ 100 mV vs. Ag/AgCl) and electrochemically reduces hydrogen peroxide generated
by a post-column IMER.

Chromatograms of (a) Ringer's solution and (b) 0.5 nmole lactate using (left) platinum electrode and (right)
wired peroxidase electrode.

Chromatograms of 1 nmole glucose (G), 0.23 nmole ascorbic acid (AA) and 0.24 nmole uric acid (UA) on a bare
platinum electrode (left). The use of the wired peroxidase electrode (right) effectively increases response.

Redox cycles for an oxidase IMER and wired peroxidase electrode. The peroxidase is covalently linked to a
redox hydrogel coated on the surface of a glassy carbon electrode. The oxidized form of the enzyme can be electroreduced through
redox centers in the polymer network.
The Kit
The wired enzyme electrode is offered in kit form. The kit provides a means of preparing and regenerating the electrode for
minimal cost. Kits are available in two formats, according to the style of BASi electrochemical detector being used. Customers
already using a BASi applications kit based on an oxidase IMER will find it very easy to upgrade to the new wired enzyme electrode
format.

Chromatograms of 0.5 pmole acetylcholine and choline with (left) platinum electrode and (right) wired
peroxidase electrode.
Contents
- Peroxidase redox polymer coating solution
- Two working electrodes (cross-flow or radial-flow)
- 1 µL dispensing syringe
- Surfactant solution
- Thin-layer cell gaskets
- Storage base and cover
- Cold packs
- Manual
Caution
Peroxidase Redox Polymer solution must be kept cold. Kits are shipped worldwide by express courier. Solution must be
refrigerated immediately upon receipt.
UniJet technology is the property of Bioanalytical Systems, Inc. US Patent 5,399,256.
Wired Enzyme sensor technology is liscenced from E. Heller and Co. US Patents 5,320,725 and 5,262,035
|