BASi Low Volume Cell for C-2 Cell Stand
Most common cells for electrochemical
analysis require sample volumes of 5-10 mL. When rare or expensive materials are studied, a lower volume is desirable. A low-volume
sample cell is available as an option for the C-2 Cell Stand.* It holds a minimum cell volume of 200 µL. The sample tube
accommodates standard BASi voltammetry electrodes. It also allows a 1/16" Teflon tube to be inserted adjacent to the working
electrode so that nitrogen can be blanketed over the sample to exclude oxygen. The normal C-2 auxiliary electrode may be placed in
the supporting electrolyte via the cell plug port. This will suffice when the solution is very conductive and the generated current
is low. The porous Vycor frit at the end of the sample chamber may introduce significant iR drop, particularly with organic
solvents or high current. In these instances, we recommend that the auxiliary electrode be placed within the working electrode
compartment as shown.
The porous Vycor frit in the sample compartment is shipped dry and must be soaked in the supporting electrolyte solution for at
least one hour prior to use. This procedure will ensure that the frit is fully wetted and has low resistance. The frit cannot be
allowed to dry out once wetted. It should be stored in supporting electrolyte solution between experiments. If it does dry, it can
be replaced (p/n MF-2064).
The C2 Low Volume Cell Kit includes: cell vial, cell top, sample chamber and o-ring, reference electrode, auxiliary electrode,
port plug, teflon gas line, and replacement vycor frits.
*Use this cell only with a C-2 Cell Stand. The cell arm of the C-2 is required to secure this cell top.
|